| ELECTRONIC REGIONAL TEST LABORATORY (WEST) | |||
|
INTRODUCTION Electronics Regional Test Laboratory (West), Mumbai, ERTL (W), Was set-up under Standardization, Testing & Quality Certification (STQC) Directorate, Ministry of Information Technology, Govt. of India, in 1985 as one of the prime national institute to satisfy the needs of Indian industries for the purpose of Test, Calibration, Training & Quality Improvement services. The laboratory presently situated at MIDC Industrial area, Andheri (E),Mumbai is equipped with sophisticated instruments and supported by a group of dedicated, professionally qualified & trained scientists / engineers. The laboratory has built-in quality system in line with ISO / IEC-17025 & NABL-101. It has received a number of recognition, both national and international under various accreditation schemes. The laboratory services have enable customers to improve their product quality and reliability & to enhance their domestic and export market share. Client range varies from typical small- scale industries to multinationals. Though, primary objective of the laboratory is to serve industries from western region, units from all over the country have been availing of specialized services of the laboratory. The transparent working system of the laboratory has made it possible for the industries to have simpler, quicker & effective interaction. |
|||
|
ERTL
|
|||
|
No. |
Parameter |
Range |
Measurement capability |
|
1. |
DC Voltage |
2 mV to 1000 V |
± 3.0 ppm to 0.022 % |
|
2. |
DC High Voltage |
Upto 10 kV |
± 0.02 % |
|
3. |
AC Voltage |
1 mV to 1000 V/ |
± 0.005 % to |
|
4. |
AC High Voltage |
1 kV to 28 kV / 50 Hz |
± 0.5 % |
|
5 |
DC Current |
2 µA to 2200 A |
± 0.2 % to 26 ppm |
|
6. |
AC Current |
10 µA to 30 A / 1kHz |
± 0.01 % |
|
7. |
DC Resistance |
0.1
Ω to 1 Ω 10 Ω to 100 MΩ 100 MΩ to 2 T Ω |
± 20 ppm to 0.06 % |
|
8. |
AC resistance |
100 µΩ to 100 MΩ / 1 kHz |
± 0.02 % to 10 % |
|
9. |
Inductance |
100 µH to 10 H |
± 0.05 % to 0.3 % |
|
10. |
Capacitance |
10 pF to 100 mF |
± 0.02 % to 1 % |
|
11. |
Frequency |
1 Hz to 18 GHz |
± 5X10E-11 to 3X10E-8 |
|
12. |
Level |
- 90 dBm to + 20 dBm/ |
± 0.15 dB to 1 dB ± 0.4 dB to 1.5 dB |
|
13. |
Power/ Energy |
I = upto 24 A |
± 0.02 % |
|
14. |
Power Factor |
0 to 1 lead & lag |
± 0.02° |
|
15. |
RPM |
4 to 99999 rpm |
± 0.008% ± 1 to 2 dgt. |
|
|||
Temperature:
|
|||
|
Sr.No. |
Parameter |
Range |
Measurement Capability |
|
1. |
Temperature |
-189 to 660°C |
0.01°C at 660°C |
|
|||
|
Sr. No. |
Parameter |
Range |
Measurement capability |
|
1 |
Temperature |
-80 to 250°C |
± 0.01°C to 0.05°C |
| *Black Body **Triple point of water |
|||
|
|||
|
Sr. No. |
Parameter |
Range |
Measurement Capability |
|
1. |
Resistance |
0 to 25 Ω |
± 0.0005 Ω |
|
2. |
Voltage |
0 to 50 mV |
± 0.004 mV |
|
3. |
Temperature |
10-1700°C |
± 0.25% rd + 1 dgt. |
| Pressure | |||
|
Sr.No. |
Parameter |
Range |
Measurement Capability |
|
1. |
Pressure |
1 to 1200 bar |
± 0.01% rd. |
|
Relative Humidity |
|||
|
Sr. No. |
Parameter |
Range |
Measurement Capability |
|
1. |
Humidity |
1 to 100% RH |
± 1% RH |
| Dimensional Metrology | |||
|
Sr. No. |
Parameter |
Range |
Measurement Capability |
|
Length measurement for following instrument/gauges |
|||
|
1 |
Vernier caliper |
0-600 mm |
± 0.8 µm |
|
2. |
Outside Micrometer |
0-500 mm |
± 0.8 µm |
|
3. |
Standard Rods |
Upto 300 mm |
± 0.4 µm |
|
4. |
Inside and depth Micrometer |
Upto 300 mm of total length (Basic Micrometer unit+ Extension piece) |
± 0.4 µm |
|
5. |
Dial Gauge |
LC upto 1µm |
± 0.4 µm |
|
6. |
Plug Gauge |
0-300 mm |
± 0.4 µm |
|
7 |
Ring Gauge |
0-200 mm |
± 0.4 µm |
|
8. |
LVDT probe |
LC upto 1 µm |
± 0.4 µm |
|
9. |
Slip Gauge |
Upto Grade ‘O’ |
± 0.4 µm |
|
10. |
Metric Scale |
Upto 1000 mm |
± 0.8 µm |
|
For further details please mail to webmaster@instrumentationguide.com |
|||