NON-ELECTRICAL ,CALIBRATION FACILITIES AT ERTL (N), NEW DELHI |
||||||
|
SR. NO. |
PARAMETER |
|
EQUIPMENT |
RANGE |
ACCURACY |
APPLICATION FOR CALIBRATION OF |
|
1 |
PRESSURE |
1 |
MANOMETER (DIGITRON P1600A) |
0 to 1900mbar (absolute) Medium Air |
+/-0.056% RDG
|
vacuum gauge compound gauge low pressure calibrator |
|
|
|
2 |
Digital Pressure Calibrator (Budenberg E400) |
0 to 400 BAR (Gauge) Medium: Water |
+/- 0.05% FS |
Industrial Pressure Gauge Pressure Transducers Pressure Switches |
|
|
|
3 |
Dead Weight Tester (Bundenberg 580 DXA) |
0 to 700 BAR (Gauge) Medium : Oil |
+/- 0.02 to 0.008% RDG |
Pressure Calibrators Reference Pressure Gauges |
|
|
|
4 |
Reference Pressure Gauge (Ravika Rav*51) |
0 to 7 kg/cm²(Gauge) Medium Air |
+/- 0.02 kg/cm² |
Low Pressure Calibrators Pressure Gauge (with restriction on air medium only) |
|
2 |
TEMPERATURE |
1 |
Oil Baths (FP 80 & SC 12 HT)with measurement system(Fluke 8520A &mRosemount RTD) |
-30 to 200 Deg C |
+/-0.025 to 0.065 Deg C |
Thermometer/ Dial gauges Sensors. Temperature Indicators with sensors Temperature controller, switches |
|
|
|
2 |
High Temperature Measurement Systems(Hart 9112 and measuring modules) |
300 to 600 Deg C 300 to 1100 Deg C
|
+/-0.475 Deg C +/-1.09 deg C |
1)RTD's Lenghts:200 mm or more Diameter:7mm, 15mm 2)Thermocouples'J','K','R','S' etc. Lenghts:200 mm or more Diameter: 7mm, 15mm |
|
3 |
HUMIDITY |
|
Digital Psychrometer(Thunder Scientific 5A-1MP) |
0to 100% RH 0 to 100 Deg C |
+/-1%RH +/- 0.3Deg C |
1) Environment Chambers (Hot/ cold/Damp) 2)Ovens 3)Incubators |
|
4 |
DIMENSION |
1 |
Slip Gauge Set and Accessories (Mitutoyo and Micronix) |
0 to 200 mm |
'O' Grade |
Vernier Calipers Micrometer Height Gauge |
|
|
|
2 |
Dial Gauge Calibrator |
0 to 25 mm |
+/- 0.002 mm |
1 mm, 5mm, 20 mm,25mm range and 0.01 mm resolution 1)Dial Indicators 2)Dial Test Indicators 3)Back Plunger Dial Indicators Linear Gauges (Probes) with indicators 5)Before Gagues(vertical and Horizontal) 6) Lever Type Gauges |
|
5 |
Weight |
1 |
Dead Weights (Mettler) |
-1 mg to 200 gm |
+/-0.008mg to +/-0.3mg (E2 Grade) |
Electronic Balance of Capacity 200 gms |
|
|
|
2 |
Dead Weights (Mettler) |
500 gm 1,2,5 & 10 kg
|
+/-0.75 mg to+/-15mg (E2 Grade) |
Electronic Balances of capacity 20 kg |
|
|
ELECTRONIC BALANCES |
1 |
Mettler (Mettler AT 201) |
205 gms/0.01mg |
+/-0.03 mg to 0.12 mg |
Weights upto 200 mgs |
|
|
|
2 |
Mettler (Mettler PG 12001-S) |
12000gms/0.1gm |
+/-0.1gm |
Weights upto 12000 gms |
|
6 |
RPM |
|
Stroboscope and Tachometer |
upto 6000 RPM |
+/- 1RPM upto 5000 RPM +/-0.05% Above 5000 RPM |
Stroboscope Tachometer (Non Contact Type) |
|
7 |
Vibration |
|
(Vibration meter IMV*) |
Acceleration upto 40g Velocity upto 500 rhm/s Displacement upto 1000 micron (all parameters at 80 Hz) |
For g +/-3% RDG For Velocity & Disp +/-5% RDG |
1) Vibration Meters 2) Vibration Systems 3) Bump Machines |
|
8 |
OTHER ITEMS |
1)Soldering Iron 2)Soldering Iron Calibration 3)Wave Soldering M/C 4) Multi Function T/C, RTD Calibrators 5)Dry Block Furnace |
||||
| Back | ||||||